網(wǎng)站首頁>技術(shù)文章 > NDJ-5S-8S-9SO Digital Viscometer perating Instruction

NDJ-5S-8S-9SO Digital Viscometer perating Instruction

Operating Instruction

-NDJ-5S-8S-9S

Digital Viscometer

CONTENT

1. Operating principle and Using …………………1

2. Main Technical Data ……………………………1

3. Application Environment ………………………2

4. Structher and Installation ………………………2

5. Using……………………………………………3

6. Panel operation…………………………………3

7. Precaution………………………………………6

8. Packing list………………………………………8  
1Operating priciple and Using

   NDJ Series Digital Viscometer (NDJ-5S-8S-9S) is a smart instrument,depending on the technology of A 16-bit high-performance single chip,which has totally changed the previous rotor method. Stepper motor runs accurately and stably as the setting speed procedure. And then torque sensor will drive rotor rotation in a constant speed. When rotor suffers viscous resistance in the tested liquid, the viscosity of the tested liquid will show on the screen by testing and handling.

   Compared with the same instrument, this one has the merit of high measure accuracy, stabled in rotate speed, wide working voltage (available in 110V, 220V, 50Hz, 60Hz) and so on. It is very convenient for users to choose the rotor and rotor velocity of the testing liquid correctly, because of the function of measured value occupies full range.

    It can widely used to measure Dynamic Viscosity of solvent based adhesive, latex, biochemicals, oil paint, coating, cosmetic, printing ink, pulp,starch, food and so on.

2: Main Technical Data

1NDJ-1S,NDJ-5S

    Measurement range10mPa.S-100,000mPa.S(1mPa.S=1cp)

    Measurement errorNDJ-1S:±5%NDJ-5S:±3%,

    Rotor types1#, 2#, 3# and 4# rotors(0# is optional accessory)

    Rotor velocity6、12、30、60RPM/min

    Dimensions95*130*155 (not including classis)

    Net weight2kg  (not including classis)

2NDJ-8S

    Measurement range10mPa.S-2,000,000mPa.S(1mPa.S=1cp)

    Measurement error:±3%

    Rotor types1#, 2#, 3# and 4# rotors(0# is optional accessory)

    Rotor velocity0.3、 0.6、1.5、3、6、1230、60RPM/min

    Dimensions95*130*155 (not including classis)

    Net weight2kg(not including classis)

3NDJ-9S

    Measurement range10mPa.S-6,000,000mPa.S(1mPa.S=1cp)

    Measurement error:±3%

    Rotor types1#, 2#, 3# and 4# rotors(0# is optional accessory)

    Measured Value will become larger as the rotor number grows

    Rotor velocity0.1、0.3、0.6、1.53、612、3060RPM/min

    Dimensions95*130*155(not including classis)

    Net weight2kg(not including classis)

3:Application Environment:

Environment temperature 5- 35

    Relative humidity:≤80%

    Power supply 220V,50Hz;110V60Hz

    There is no electro magnetic interference, severe virbation, corrosive gas around the instrument.


4:Structure and Installation

 1Structure

            As follows:

1level babble  2screen  3cover  4rotor protection frame

5the host base 6operation key 7rotor connector  

8rotor9the host base horizontal knob

2Installation

1. You need to take classis, stanchion, lifting block and hand shank from the packing box. Screwing stanchions into the hole on the base. (The rack side must face to you), tiding the hand shank of viscomeler to the lifting block.

2. Adjust the tightness of up and down by swirling the up-down button, then putT"block behind instrument into hand shank and screw them.

3. Adjust three level screws of the base to make the level bubble in the center of  black circle.

4. Take off universal-joint protective cap under the main case.

5. Connect power

5:Operation

1. Prepare the liquid to be measured and put it into a round fat container or flat container with the diameter not smaller than 60mm, you need to control temperature of measured liquid correctly (It will influence viscometer directly)

2. Please reverse screw protection bracket beneath the main case.

3. Please choose the right rotor and contrarotate in the universal joint ofthe instrument

4. Adjust the lifting screw and put the rotor into the measured liquid till the level mark on the rotor reach the liquid surface.

5. Adjust the instrument in level again.

6. During the testing, the environment temperature must be stable to make sure the show value is accurate.


6Panel operation

This instrument has eight keys: choosing rotor, setting rotor speed,,,

stat, stop,search, confirm. Turn on the power switch, enter standby state.

7Precaution

1. Because Tem is viscosity's function, so the Tem measurement error should  be  controlled in ±0.1℃  when  the  instrument works in the normal temperaure.

2. Please look out the aperture angle percentage, its value should be kept 20%-85% . When the valuce  is  too  high  or  too  low,  please change rotor or velocity. Otherwise, it will affect measurement accuracy.

3. The instrument should be used in the required range of voltage  and frequency, Othervise, it will afect measurement accuracy.

4. You should operate carefully when you put on or take off rotor.  You just need to pull up the universal joints a little bit, forbidding  pushing too  hard.

5. Afer using, universal joints and rotor should be kept clean.

6. You need to drop the instrument slowly. You'd better hold it to  avoid damage the  axis because of vibration.

7. Universal joints should be put on protctive nap when the  instrument is moved or  transported.

8. Turbid liquid, emulsion, high polymer material and some high viscosity liquid, which all  belong non-Newtonian liquid. Their apparent viscosity and velocity are not linear relationship. So,  when  you  use  different rotors, velocity, and measurement time, their viscosity measurement may be different, not the mistake of instrument testing.

9. You will get accurate result when you do as fllowing:

    A. Control the tested liquid temperature accurately.

    B. Kcep environment temperature balance.

    C. Rotor and the tested liquid must have the same temperature.

    D. The bigger rotor, higher speed usually is used for low viscometer

     liquid, and high viscometer usually uscs smaller, lower speed.

    E. Keep the surface of rotor clean.

Attach: the installation step of  0#

1): The using of 0# must be followed as instument operations. You will    know the specification by picture.

2): Pouring 25-30mL tested liquid into sleeve which has bottom.

3): Put on the out sleeve from the bottom up and fix it. (Sleeve P9  has V-groove)

4): The same way to use sleeve. When the  out  sleeve  and  rotor  are immersed in liquid, taking the out sleeve  having  V-groove  inside as the liquid line of  0#.

Packing  list

No.           Name                               Number

1      NDJ series Digital Rotary Viscometer                 1

2      Stanchion and lifting block                          1

3     Protection bracket                               1

4      Rotors, 1#, 2#, 3# and 4#                          1

5     Power line                                      1

6      Operation manual                                1

7     Quality certificate                                1

8      Guarantee card                                  1

8Attach

Installation step:

Screw 0# rotor (5); fixed tap of sleeve (2); Screw in screw (1) and fix it;

Put on size zero sleeve (4); Screw in screw (1) and fix it.



TEL:13798128437

掃碼添加微信
国产乱子夫妻xx黑人xyx真爽| 人妻体内射精一区二区三四| 中文亚洲av片在线观看| 小辣椒福利视频精品导航| 丰满熟妇乱又伦在线无码视频| 久久综合一区二区三区| 四虎成人精品无码| 又粗又猛又黄又爽无遮挡| 精品国产乱码久久久久久婷婷| 国产啪精品视频网站| 亚洲成aⅴ人最新无码| 国产一在线精品一区在线观看| 东京热加勒比日韩精品| 麻豆成人久久精品二区三区免费| 日本中文字幕有码在线视频| 69sex久久精品国产麻豆| 97影院理论午夜伦不卡| 欧美另类高清zo欧美| 亚洲制服丝袜一区二区三区| 久久久久久免费毛片精品| 国产精品午夜小视频观看| 日本在线一级高清自拍| 久久婷婷国产综合精品| 国产一区二区内部视频| 精品无码国产污污污免费网站国产| 久久精品噜噜噜成人| 亚洲av成人专区精品| 色噜噜狠狠一区二区三区| 少妇泬出白浆18p| 日日碰狠狠躁久久躁综合网| 丰满妇女毛茸茸刮毛| 狠狠88综合久久久久综合网| 野花香社区在线视频观看播放| 亚洲中文字幕国产剧情| 性色av一区二区三区| 欧美丰满一区二区免费视频| 日韩精品亚洲一区二区三区| 中文字幕精品久久久乱码乱码 | 久久精品国产清高在天天线| 免费a级毛片无码a∨| 国产9 9在线 | 中文|